The following table lists the changes that have been made to the
CVE-2025-1204 vulnerability over time.
Vulnerability history details can be useful for understanding the evolution
of a vulnerability, and for identifying the most recent changes that may
impact the vulnerability’s severity, exploitability, or other characteristics.
-
New CVE Received
by [email protected]Feb. 25, 2025
Action Type Old Value New Value Added Description The “update” binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the ‘C’ button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device. Added CVSS V4.0 AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X Added CWE CWE-912 Added Reference https://claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a-chinese-backdoor-the-reality-is-more-complicated?ref=vault33.org Added Reference https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01